Line 42: Line 42:
 ^121GW SN906 | Penge | 4.9998 V | 4.9931 V | 5.0224 V | 0.9945 mA | 0.9947 mA | 100.11 | 0.9996 | 9.999 | 100.006 | 999.97 Hz | 9.9998 Hz | ^121GW SN906 | Penge | 4.9998 V | 4.9931 V | 5.0224 V | 0.9945 mA | 0.9947 mA | 100.11 | 0.9996 | 9.999 | 100.006 | 999.97 Hz | 9.9998 Hz |
 ^121GW SN907 | Penge | 5.0004 V | 5.0006 V | 5.0808V | 0.9703 mA | 1.0011 mA | 0.9692 mA  | 100.10 | 0.9998 | 10.000 | 100.06 | 100.00 | 10.000 | ^121GW SN907 | Penge | 5.0004 V | 5.0006 V | 5.0808V | 0.9703 mA | 1.0011 mA | 0.9692 mA  | 100.10 | 0.9998 | 10.000 | 100.06 | 100.00 | 10.000 |
-^TTI 1906 Computing Multimeter | Penge | 4.9992 V | 5.9970 V | 4.9634 V | 1.00008 mA | 1.00010 mA | 1.00012 mA | 100.155 | 0.99988 | 10.0023 | 10.0023 | 100.090 |+^TTI 1906 Computing Multimeter | Penge | 4.9995 V | 5.9970 V | 4.9634 V | 1.00008 mA | 1.00010 mA | 1.00012 mA | 100.155 | 0.99988 | 10.0023 | 10.0023 | 100.090 
 +^Keithley 2750 |1000|5.00005v|4.99542v|4.99180v|1.00015mA|0.001287A|0.001190A|100.0596Ω|0.999841Ω|10.00276kΩ|100.1022kΩ|100.0042Hz|10.00042kHz|
  
Permalink equipment/instruments/dmmcheck.1645721616.txt.gz · Last modified: 2022/02/24 17:53 by Georg Sluyterman

oeffentlich